Reliability of transcutaneous bilirubin estimation during phototherapy in term infants: A hospital-based observational study
Jehangir Allam Bhat1, Sajad Ahmad Sheikh2, Zubair Ahmad Wani3, Roshan Ara4, Sami Ullah Bhat5
1 Department of Paediatrics, Vikas Hospital Pvt. Ltd, New Delhi, India 2 Department of Gastroenterology, Vikas Hospital Pvt. Ltd, New Delhi, India 3 Department of Paediatrics, Government Medical College, Srinagar, Jammu and Kashmir, India 4 Department of Accident and Emergency, Vikas Hospital Pvt. Ltd, New Delhi, India 5 School of Studies in Chemistry, Jiwaji University, Gwalior, Madhya Pradesh, India
Correspondence Address:
Dr. Jehangir Allam Bhat Department of Paediatrics, Kurji Holy Family Hospital, Patna, Bihar India
Source of Support: None, Conflict of Interest: None | Check |
DOI: 10.4103/smj.smj_11_19
|
|
Background:Blood sampling method is the most common but invasive method for bilirubin estimation. Transcutaneous bilirubin (TCB) estimation, on the other hand is non-invasive. Objectives: The aim was to compare neonatal bilirubin estimation by serum bilirubin and transcutaneous bilirubinometer methods both pre- and during phototherapy and establish the precision of transcutaneous bilirubinometer. Materials and Methods: This prospective hospital-based observational study was conducted on 350 term neonates over a period of 2 years. Bilirubin was estimated in all newborn by both serum bilirubin and transcutaneous method, at the time of admission in nursery (prephototherapy), at 6 h after phototherapy (while neonates were still under phototherapy treatment), and at withholding phototherapy, and a photo opaque patch of 2.5 cm was used to cover the area on midforehead and transcutaneous bilirubin was estimated from this area after removing it during phototherapy. All data were tabulated and analyzed by relevant statistical tests. Results: A total of 350 term healthy newborns with mean gestational age of 38 ± 2 weeks and mean birth weight of 2.97 ± 0.9 kg were studied. Correlation analysis revealed a significant correlation between total serum bilirubin and transcutaneous method (TCB) measurements of bilirubin. When these methods were compared at pre-phototherapy, at 6 hours after phototherapy and before withholding phototherapy, correlation coefficient(r) of 0.9571, 0.7988, and 0.9206 were seen respectively. The precision levels of transcutaneous bilirubinometer (BiliCheck™) were good because plotting of Bland–Altman graphs showed maximum (>97%) data within agreement limits at all stages of phototherapy. Conclusion: Our study revealed good correlation and precision of transcutaneous bilirubinometer both before and during phototherapy. Thus, it supports the previous studies which justify the use of transcutaneous bilirubinometer in assessing newborn jaundice during phototherapy.
|
|
|
|
[FULL TEXT] [PDF]* |
|
|
|